40 kW Two-Pulse diagnostic X-ray generator
The Heliophos D X-ray generator can be used for routine radiographic
and fluoroscopic diagnostic procedures of the skeletal system, respiratory system,
obstetric pelvis, spinal column and gastro intestinal tract.
• Radiographic output: 40kW as per IS:7620 (Part 2)
• Tube filament current stabilisation through frequency insensitive
voltage stabiliser
• Full wave rectification
• Digital display of kV, mA and mAs
• Overload protection (Blinking of display and blocking
of exposure)
• Shortest exposure time: 10ms
• Supports serialography upto 2 fps
• kV, mA and s values set with step switches as per
Siemens point technique
• High kV radiography
• Fluoroscopy time totaliser
• Rapid braking of rotating anode increasing tube life
• Supports upto 5 examination units
Technical Description
Heavy-duty tube units can be connected. The register
setting techniques ensures tube overload protection and simple operation. The
precision electronic timer and contactor combination are
Heliophos D
40 kW Two-pulse diagnostic X-ray generator with Digital
Display
engineered for “zero-phased” radiography exposures. Even
at high frame rates, like in serialography, voltage spikes and transients that
may affect the X-ray tube are eliminated.
Salient features
• Radiographic output: 40kW as per IS:7620 (Part 2)
• Tube filament current stabilisation through frequency insensitive
voltage stabiliser
• Full wave rectification
• Digital display of kV, mA and mAs
• Overload protection (Blinking of display and blocking
of exposure)
• Shortest exposure time: 10ms
• Supports serialography upto 2 fps
• kV, mA and s values set with step switches as per
Siemens point technique
• High kV radiography
• Fluoroscopy time totaliser
• Rapid braking of rotating anode increasing tube life
• Supports upto 5 examination units
Technical Description
Heavy-duty tube units can be connected. The register
setting techniques ensures tube overload protection and simple operation. The
precision electronic timer and contactor combination are
Heliophos D
40 kW Two-pulse diagnostic X-ray generator with Digital
Display
engineered for “zero-phased” radiography exposures. Even
at high frame rates, like in serialography, voltage spikes and transients that
may affect the X-ray tube are eliminated.
Heliophos D
Technical Data at a glance
Control console and High Tension
Transformer
Output power: 40kW as per IS:7620-1986 (Part 2) 500 mA at 70 kV; 320 mA at 90 kV (with THX1 Tube) 500 mA at 100 kV; 320 mA at 125 kV (with THX2 Tube)
Radiography parameters: kV
range: 36 to 125 kV in 25 steps
mA range: Small
focus (80, 100,160); Large focus (160, 320, 500)
mAs range: Small
focus (1-500); Large focus (1.5 - 192)
Exposure time: min
10 ms to maximum 5 sec in 25 steps Digital display of Radiographic Parameters – kV, mA &
mAs
Fluoroscopy parameters: kV
range: 40 to 90 kV in 11 steps mA range: 0-3 mA
Fluoroscopy timing: Set
by time totaliser, maximum time 5 minutes, alarm after elapse of set timing
Radiography technique: Three
point technique
Exposure release: From
control console with two step switch or from examination unit
Fluoroscopy technique: independent
adjustment of kV and mA from control console or remote control from examination unit
Fluoroscopy release: From
control console or from examination unit
Mains Voltage compensation: Provided (354 - 447 Volts)
Filament stabilisation: Provided
Anode braking device: Provided
Overload Protection: Provided,
blocking for any invalid combination
Serialography support: Yes,
upto 2 frames per second
Examination units: Upto
5
X-ray tubes: Upto
2
X-ray tube
Type of tube: Rotating
Anode X-ray tube BEL THX1 - 125 / 20 / 40;
BEL THX2 - 125 / 30 / 50
Focal spot: Small
focus: 1.2 x 1.2 mm
Large focus: 2.0
x 2.0 mm
Anode speed: 2800
rpm
Anode HU: 115000
(THX 1); 260000 (THX 2)
Inherent filteration: 2 mm Al
HT connection Through
pair of High Tension Cables (Standard length 8 metres)
Weight of assembly: 18
kg (THX 1); 25 kg (THX 2)
Installation information
Line connection 415
V + 10%, -15%, 50Hz, 63 A, mains resistance ≤ 0.36 ohms
Power consumption 50
kVA
Dimension: control console: 672 x 395 x 895 mm
HT Transformer: 426
x 365 x 800 mm
Weight: control console: 140Kg
HT transformer: 148
Kg
Cable length from control to HT trafo: 5.5 metres Temperature
range +10° to 40° C Relative humidity 20% to 75 % non condensing
Barometric pressure 700
to 1060 HPa
Safety standards IS: 7620
and AERB Safety Code
At Siemens we are fully committed to radiation safety.
The guiding spirit behind the factory in Goa is to provide
high quality world class medical equipment at affordable price.
A wide range of X-ray systems - mobile
units, radiography systems and fluoroscopy systems – and ultrasound
equipment
are manufactured here. Outside Europe, the factory in Goa is one of the largest
manufacturing base
for X-ray products. Within a short period from its
inception, the factory has been awarded the seal of quality - the
ISO 9001 :
1994 certification - by the Underwriter Laboratories (UL) of USA. The factory
is also certified by the
Bureau of Indian Standards for IS:7620 (Part 1). The
radiation safety of products manufactured in Goa are certified
by the Atomic
Energy Regulatory Board (AERB), Government of India. The high degree of quality
consciousness
drives the factory towards higher levels of quality standards
like Good Manufacturing Practices (GMP) of USA.
..... backed by the efficient Siemens UPTIME Service
What UPTIME means to you
• Round the clock call handling, all the seven days of
a week.
• Immediate response
• Faster problem resolution
• Faster parts determination and coordination
• Increased effectiveness of your on-site engineer
• Pro-active service
• Latest information on the status of the call as and
when you require. Call 9622 000 007 or log on to
www.service007.com
• Enhanced applications support
• Flexibility to choose from a range of “modular”
service agreements
Siemens Ltd.
MED-41-383-004
(This replaces MED-41-383-002)
Product development is a continuous process.Consequently the data indicated in this brochure is subject to
change without prior notice.
Head Office
Siemens Ltd.
Medical Solutions Division,
130, Pandurang Budhkar Marg,
Worli, Mumbai 400 018.
Tel.: 022 498 7893
Fax: 022 493 0083
Visit us at: www.siemensmedical.com
For futher details please contact:
Chennai Tel: 044 823 1125/26, Fax: 044 827 5660
Kolkata Tel: 033 442 8471/72//76, Fax: 033 442 1147
Mumbai Tel: 022 495 1071 / 494 1884, Fax: 022 496 3735
New Delhi Tel: 011 332 6722, Fax: 011 332 6147
----------------------------------------------------------------------------------------------
'Product development is a continuous process. Consequently the data indicated in this brochure is subject to
change without prior notice.
Siemens Ltd.
MED-41-383-004
(This replaces MED-41-383-002)